A2) patient date of birth - not available from facility.A3b) patient gender - not available from facility.A4) patient weight - not available from facility.A5) patient ethnicity - not available from facility.B6/b7) patient information regarding relevant tests/laboratory data or medical history - not available from facility.H3) a portion of the device was discarded, and a portion remained within the patient, thus no product investigation could be performed.H6) although lld cut/cap is a known risk of complication with use of the lld, the physician did not attempt to unlock the lld from the lead prior to cutting and capping.Submission of this report does not, in itself, represent a conclusion by the manufacturer and/or authorized representative or the national competent authority that the content of this report is complete or accurate, that the medical device(s) listed failed in any manner and/or that the medical device(s) caused or contributed to an alleged death or deterioration in the state of the health of any person.
|
|
A lead extraction procedure commenced to remove a right atrial (ra) and a right ventricular (rv) lead due to redundancy.Spectranetics lld #2 lead locking devices (lld #2s) were inserted into each lead to provide traction.Beginning with a spectranetics glidelight laser sheath on both leads, little progress was made due to heavy calcification.Switching to a spectranetics 13f tightrail rotating dilator sheath on the ra lead, advancement was made through the superior vena cava (svc).A pleural effusion was detected via venogram, and a subclavian perforation was noted (mdr #3007284006-2025-00077).Rescue efforts began, and the tightrail device was used to stop bleeding from the perforation.A rescue balloon was used to tamponade the injury, and a chest drain was placed to remove the blood.The patient stabilized, and venogram showed no effusion, but minimal blood loss was coming from the chest tube.As a result, the decision was made to stop the extraction.No attempt was made to unlock the llds from the rv lead, (mdr #3007284006-2025-00078), or the ra lead (mdr #3007284006-2025-00079), so the leads, along with the llds, were cut and capped and remained in the patient.The patient survived the procedure.This report captures the lld #2 within the ra lead, which was cut/capped and remained in the patient.There was no alleged malfunction of any spectranetics devices in use during the procedure.
|