| Model Number |
RELTACK4XDPT |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/08/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information: b5, g3.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that during the laparoscopic inguinal hernia repair , during mesh fixation, the reload disengaged into the patient and was subsequently removed with a laparoscopic grasper.Another tacker was opened and used to complete the procedure.No patient com plications have been reported as a result of this event.
|
| |
|
Event or Problem Description
|
|
It was reported that during the laparoscopic inguinal hernia repair , during mesh fixation, the tip of the tacker fell into the patient and was subsequently removed with a laparoscopic grasper.Additionally, the reload disengaged also into the cavity of the patient.Another tacker was opened and used to complete the procedure.No patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction: g1.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: b5, g3, h3, h6 h3.Evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Functional testing found that the articulation knob functioned properly.The sulu was reloaded.The handle could be actuated and cycled properly and the next tack was protruding.A test sulu from lot #n4h1907y could be loaded onto the returned instrument.The instrument and test sulu were applied to the appropriate test media.The tacks deployed and seated properly in the test media.It was reported that the sulu was disengaged.The reported issue was confirmed.The product analysis noted evidence that the device was not used as intended.This issue could be due to excessive manipulation or to improper loading and/or unloading of the sulu.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: caution: do not attempt to remove the shipping wedge until the reload is locked and loaded onto the instrument.Ensure the device is in reload mode, red toggle button exposed, and the device is in the straight (unarticulated) position.To load the reliatack¿ device select the appropriate reload, then pull and hold the blue lock switch button, located on top of the instrument, in the back position.Insert the pin located at the distal end of the shaft into the reload.Ensure that the arrow on the shaft is aligned with the arrow on the reload.Push the reload onto the distal end of the shaft until the two arrows meet.Release the lock switch button allowing it to return to the forward position.Gently pull on the reload to confirm that the device is properly loaded.Remove the shipping wedge once the reload is properly locked and loaded.Instructions for unloading: push the left side of the toggle button to expose the red portion of the button and ensure that the device is in the straight (unarticulated) position.Pull and hold the blue lock switch button, located on top of the instrument, in the back position.Pull the reload off of the shaft of the device.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that during the laparoscopic inguinal hernia repair , during mesh fixation, the reload disengaged into the patient and was subsequently removed with a laparoscopic grasper.Another tacker was opened and used to complete the procedure.No patient com plications have been reported as a result of this event.
|
| |
|