Catalog Number 101/810/080
Medical Device Problem Code Gas/Air Leak (2946)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/01/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that there were issues with two cannulas.The first was changed and its cuff lost air during the night, requiring replacement the next day.The patient continued treatment without issue for four days before the second cannula also experienced air leakage, requiring another replacement.There was patient involvement and no harm reported.
 
Additional Manufacturer Narrative
B3: day of event unknown h3: the device history record of reported lot was reviewed and it was confirmed that no non-conformities were reported during production.The quality inspection forms were reviewed, and it was observed that no defects were found, and the lot was released.The complaint for the failure mode could not be confirmed by investigation as no samples nor pictures was provided by the customer.Without the return of the samples a comprehensive failure investigation cannot be performed and a probable cause cannot be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX TUBES BLUSELECT SUCTIONAID
Common Device Name
TUBE, TRACHEOSTOMY (W/WO CONNECTOR)
Manufacturer (Section D)
SMITHS HEALTHCARE MANUFACTURING S.A. DE C.V.
ave calidad no. 4, parque
tijuana
MX 
Manufacturer Contact
reed covert
6000 nathan lane north
minneapolis, MN 55442
2247062300
MDR Report Key21894516
Report Number3012307300-2025-04736
Device Sequence Number12802447
Product Code BTO
UDI-Device Identifier15019517077018
UDI-Public(01)15019517077018(17)291021(10)6055542
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K173384
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 04/23/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Catalogue Number101/810/080
Device Lot Number6055542
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/26/2025
Initial Report FDA Received Date04/23/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/22/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-