• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARGON MEDICAL DEVICES SUPERCORE BIOPSY INSTRUMENT 18GA X 15CM; INSTRUMENT, BIOPSY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARGON MEDICAL DEVICES SUPERCORE BIOPSY INSTRUMENT 18GA X 15CM; INSTRUMENT, BIOPSY Back to Search Results
Model Number 701118150
Medical Device Problem Code Firing Problem (4011)
Health Effect - Clinical Code Insufficient Information (4580)
Type of Reportable Event Serious Injury
Event or Problem Description
The reporter stated, "the trigger mechanism is now extremely stiff, and in some cases, tissue samples are not being retrieved properly.".
 
Additional Manufacturer Narrative
The sample is indicated as available for return.As of the date of this report, the sample has not been returned.A follow-up report will be provided once the device has been received and reviewed.
 
Additional Manufacturer Narrative
We conducted a thorough review of the manufacturing and inspection records for this lot and found no deviations or non-conformances.Unfortunately, we did not receive any samples for evaluation, nor did we receive any photographic or visual evidence that would have allowed us to review the allegation in detail.Without this necessary evidence, we are unable to perform a thorough investigation or determine a root cause and corrective action at this time.As a result, no corrective action is required at this point.However, if samples are returned at a later date, we would be happy to reopen this complaint for re-evaluation.Additional information provided in h.6.And h.11.
 
Additional Manufacturer Narrative
A comprehensive review of the manufacturing and inspection records for this lot has been completed.No deviations or non-conformances were identified.Samples were returned by the customer for examination.A visual inspection and testing was conducted on the returned product.It was observed that there was resistance during cocking and firing of the device.The complaint was confirmed.An in-depth review was initiated to identify any gaps in the process which may lead to such issues.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUPERCORE BIOPSY INSTRUMENT 18GA X 15CM
Common Device Name
INSTRUMENT, BIOPSY
Manufacturer (Section D)
ARGON MEDICAL DEVICES
1445 flat creek road
athens TX 75751
Manufacturer Contact
kelly mayer
1445 flat creek road
athens, TX 75751
9036759321
MDR Report Key21894925
Report Number0001625425-2025-00999
Device Sequence Number15522062
Product Code KNW
UDI-Device Identifier00886333005888
UDI-Public(01)00886333005888(17)290918(10)11589736
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K974814
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/11/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number701118150
Device Lot Number11589736
Was Device Available for Evaluation? Device Returned to Manufacturer
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/25/2025
Supplement Date Received by Manufacturer06/13/2025
07/31/2025
Initial Report FDA Received Date04/23/2025
Supplement Report FDA Received Date07/03/2025
08/11/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/28/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-