• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENDOLOGIX AFX2; BIFURCATED STENT GRAFT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENDOLOGIX AFX2; BIFURCATED STENT GRAFT Back to Search Results
Model Number BEA25-100/I20-40
Medical Device Problem Codes Off-Label Use (1494); Lack of Effect (4065)
Health Effect - Clinical Codes Aneurysm (1708); Clinical Endoleak (4843)
Date of Event 03/01/2025
Type of Reportable Event Serious Injury
Event or Problem Description
The patient was initially treated for an abdominal aortic aneurysm (aaa) with implant of an afx2 bifurcated stent graft and an afx vela suprarenal aortic extension on (b)(6) 2020.On an unknown date in 2022, a type 1b endoleak coming from the right iliac artery was detected.The physician elected to implant an afx iliac extension in the right iliac on (b)(6) 2022.Additionally, embolization of a lumbar artery due to a type 2 (non-device related) endoleak was performed on an unknown date in 2024.This event was previously reported under mdr number 3011063223-2025-00049.Approximately four (4) years post initial procedure, at the last computed tomography (ct) scan in (b)(6) 2025, the aneurysm diameter measured 103mm, while in 2024 the diameter it was 102mm.The physician suspected there is a type 1b endoleak from the left iliac and elected to treat the patient.On (b)(6) 2025 two (2) afx iliac limbs were implanted.It was reported that it was difficult to pass the lunderquist (non-endologix) wire through the iliac limb; therefore, cannulation was done through the brachial branch to go through the prothesis and it was successful.The two afx limbs were implanted and at the end of the procedure no endoleak was seen.The physician will perform a ct scan in one month.
 
Additional Manufacturer Narrative
The device involved in this event will not be returned for evaluation as it remains implanted in the patient.Patient medical records and imaging studies will be requested for further evaluation by the clinical specialist.If additional information pertinent to the incident is obtained, a follow-up report will be submitted.Device iteration is afx2.
 
Additional Manufacturer Narrative
The reported adverse event/incident was investigated in alignment with endologix operating procedures and work instructions.An evaluation of the manufacturing record was completed.A review of the part number/lot number combination needed for device identification shows the device to have been properly manufactured and released in accordance with the device master record.The review confirms there were no manufacturing or processing non-conformities identified that would contribute to the reported adverse event/incident.An evaluation of the device could not be completed.The device was not returned to endologix for evaluation because it remains implanted.A clinical evaluation of the adverse event/incident was completed.An examination of medical records and/or medical imaging received by endologix shows that the aneurysm enlargement and type 1b endoleak of the left common iliac artery are unconfirmed.The additional endovascular procedure is confirmed.This is moderately consistent with the reported adverse event/incident.Device, user, procedure or anatomy relatedness of complaint could not be determined.Procedure related harms could not be determined.The clinical assessment determined a type ii endoleak (non-device related) occurred that was not in the event as reported.The type ii endoleak was discovered during a review of the ct scan labeled post op.Additionally, a 25 mm suprarenal cuff was used in conjunction with a 25 mm main body afx bifurcated stent graft.According to the afx instructions for use (ifu), when selecting a 22mm, 25mm, or 28-mm proximal extension, the diameter should be one size larger than the main body to ensure adequate overlap and sealing.It is unlikely this contributed to the reported event.The final patient status was not reported.No additional investigation of this reported adverse event/incident is planned.However, should additional information relevant to the investigation outcome become available, a follow-up report will be submitted.Endologix will continue to monitor this and similar adverse events/incidents.Corrections: g3: awareness date has been updated.H6: medical device problem codes: remove code 4065.H6: investigation finding codes: remove code 3233.H6: investigation conclusion codes: remove code 11.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AFX2
Common Device Name
BIFURCATED STENT GRAFT
Manufacturer (Section D)
ENDOLOGIX
2 musick
irvine CA 92618
Manufacturer (Section G)
ENDOLOGIX
2 musick
irvine CA 92618
Manufacturer Contact
gary kirchgater
2 musick
irvine, CA 92618
8009832284
MDR Report Key21895141
Report Number3011063223-2025-00051
Device Sequence Number8162127
Product Code MIH
UDI-Device Identifier00818009014672
UDI-Public(01)00818009014672(17)220125
Combination Product (Y/N)N
Initial Reporter CountryBE
PMA/510(K) Number
P040002
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/08/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/25/2022
Device Model NumberBEA25-100/I20-40
Device Lot Number2043002R001
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/08/2025
Supplement Date Received by Manufacturer05/27/2025
Initial Report FDA Received Date04/23/2025
Supplement Report FDA Received Date06/24/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured02/12/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
AFX ILIAC LIMB, LOT # 2355436003.; AFX VELA SUPRARENAL, LOT # 2292021008.
Outcome Attributed to Adverse Event Required Intervention;
Patient Age69 YR
Patient SexUnknown
-
-