| Catalog Number |
640CX0410 |
| Medical Device Problem Code |
Separation Failure (2547)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/07/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The healthcare professional reported that during an endovascular embolization procedure, the microcatheter (competitor brand) was in place and the 2mm x 4cm orbit galaxy helical xtrasoft (640cx0410 / 31382004) was delivered to the target site.The coil filled the aneurysm, and the physician attempted to detach the coil, but the coil was unable to be detached.The physician retracted only the coil and replaced the coil to complete the procedure using the original microcatheter.There was no report of any negative impact to the patient.On 10-apr-2025, it was reported that additional information related to the reported device issue is not available.
|
| |
|
Additional Manufacturer Narrative
|
|
Manufacturer¿s ref.No: (b)(4).Information regarding patient identifier, date of birth, age, sex, gender, weight, race, and ethnicity were not provided.Section d.2b: procode is krd/hcg.Section e.1: the initial reporter phone is not available / reported.Section h.3: the device is available to be returned for evaluation and testing.However, it has not been received to date.If the device returns, a device investigation will be performed.A review of manufacturing documentation associated with this lot (31382004) presented no issues during the manufacturing or inspection processes related to the reported complaint.There were no internal actions related to device manufacture or inspection.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by johnson & johnson medtech, or its employees that the report constitutes an admission that the product, johnson & johnson medtech, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Missing information from this report is identified as blank; this information was not provided in the reported event or available at the time of report submission.The manufacturer will submit a supplemental report if new facts arise which materially alter information submitted in a previous mdr report.Additional information will be submitted within 30 days of receipt.
|
| |
|
Additional Manufacturer Narrative
|
|
Manufacturer¿s ref.No: (b)(4).The purpose of this mdr submission is to report that the product analysis lab received the complaint device on 06-may-2025.A supplemental 3500a report will be submitted once the product investigation has been completed.Updated sections: b.4, d.9, g.3, g.6.H.2, h.3, h.6, and h.11.The manufacturer will submit a supplemental report if new facts arise which materially alter information submitted in a previous mdr report.Additional information will be submitted within 30 days of receipt.
|
| |
|
Additional Manufacturer Narrative
|
|
Manufacturer¿s ref.No: (b)(4).The purpose of this mdr submission is to report the investigational finding of the returned device.The complaint product was returned and received for evaluation and analysis.The investigation is documented below.Investigation summary: a non-sterile 2mm x 4cm orbit galaxy helical xtrasoft was received contained in the decontamination pouch.Visual inspection was performed.No appearance of damage was observed.The coil was noted to still be attached to the delivery system.The embolic coil component was inspected under the microscope.It was observed to be in good normal condition (i.E.No elongations, no kinks and no non-concentric loops were noted).It was still attached to the gripper.The coil was attached to a lab sample trufill dcs syringe, then the pressure was increased in the syringe by rotating the syringe knob clockwise until the gauge indicator reaches the green zone, and the pressure rapidly decreased, indicating that the coil has been successfully detached.The functional test was performed successfully.Additionally, no damages were found on the coil that could have contributed to the issue reported during the procedure.The customer complaint cannot be confirmed with the evidence available.It is possible that other clinical and procedural factors that cannot be replicated during the analysis may have contributed to the reported failure.A review of manufacturing documentation associated with this lot (31382004) presented no issues during the manufacturing or inspection processes related to the reported complaint.There were no internal actions related to device manufacture or inspection.As part of johnson & johnson medtech quality process, all devices are manufactured, inspected, and released to approved specifications.It should be noted that product failure could be caused by multiple factors.The instruction for use (ifu) does contain the following recommendations: ¿ coil detachment must be confirmed under fluoroscopy by releasing the second rhv and slowly retracting the delivery tube ensuring that the coil is not being retracted into the tip of the infusion catheter and that the delivery tube marker bands move away from the coil without resistance.Based on the manufacturing documentation review, there is no indication that the event is related to the device manufacturing process.As part of the post market surveillance program, information from this complaint is trended for statistical signals and corrective / preventive action may be triggered at a later time.Since there was no evidence to suggest the event was related to a manufacturing or design issue, no corrective actions will be taken at this time.The manufacturer will submit a supplemental report if new facts arise which materially alter information submitted in a previous mdr report.Additional information will be submitted within 30 days of receipt.
|
| |
|
Search Alerts/Recalls
|
|