| Model Number |
VTICMO 12.6 (-17.0/2.5) |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Visual Disturbances (2140)
|
| Date of Event |
02/10/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
H6 - investigation type 4110: lens work order search - no similar complaint event(s) within associated lots were found.Secondary surgical intervention to remove/replace/reposition the lens is identified in the labeling as a known potential adverse event following icl implantation.Claim#: (b)(4).
|
| |
|
Event or Problem Description
|
|
A facility representative reported that following an implantable collamer lens (icl) implantation procedure, the patient experienced lens rotation not associated to a low vault, blurred vision, and double vision.The lens was explanted and the problem resolved.Cause of the event was reported as unknown.Reportedly, "lens rotation and decenter to inferior." there are multiple medical device reports associated with this patient.This report is 2 of 2 total reports.Based on the information provided, the risk assessment for our product, and our experience, we have determined the cause or contribution to the reported issue is not indicative of a manufacturing related issue or product deficiency.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional data: h3 - device evaluation: the lens was returned in a microcentrifuge vial with moisture and residue/debris on the product.Visual inspection found no visible damage to the lens.Dimensional inspection found the lens to be within specifications.Claim#: (b)(4).
|
| |
|