• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. POROUS TIBIA BSEPLATE W/JRNY LOCK SZ4 LT; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. POROUS TIBIA BSEPLATE W/JRNY LOCK SZ4 LT; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO Back to Search Results
Catalog Number 71425324
Medical Device Problem Code Implant subsidence (4078)
Health Effect - Clinical Codes Loss of Range of Motion (2032); Discomfort (2330)
Date of Event 04/03/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).This complaint was opened by smith+nephew to document a reported patient complication that includes reference to the use of a smith+nephew product without evidence of a specific product problem.The reported complication relates to known inherent procedural risks that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to confirm a relationship between the reported event and the device or identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are conservatively submitting this report in accordance with applicable regulations.If additional information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
 
Event or Problem Description
It was reported that, after tka surgery had been performed on (b)(6) 2024, the patient experienced sinking of the tibial baseplate and discomfort.The surgeon believed that the patient's stiff joint, had caused the stress to be concentrated in the anterior tibia, which was the cause of the sinking.This adverse event was solved by revision surgery on (b)(6) 2025, in which the baseplate was exchanged to journey tibial baseplate size 2 and genesis long stem 12mm×150mm.Patient's current health status is unknown.No further complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POROUS TIBIA BSEPLATE W/JRNY LOCK SZ4 LT
Common Device Name
PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
1450 brooks rd.
memphis, TN 38116
5123913905
MDR Report Key21896148
Report Number1020279-2025-00695
Device Sequence Number19693611
Product Code MBH
UDI-Device Identifier00885556733059
UDI-Public885556733059
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K211221
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 04/23/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number71425324
Device Lot Number21KM20129A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/03/2025
Initial Report FDA Received Date04/23/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured01/04/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
GNS II BICONVEX PAT 23MM: 71420566, (B)(6) ; JRNY II BCS FMRL OXIN LT SZ4: 74022124, (B)(6) ; JRNY II BCS XLP IST 3-4 LT9MM: 74027241, (B)(6)
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age80 YR
Patient SexMale
-
-