Model Number 5568-53
Medical Device Problem Codes Electrical /Electronic Property Problem (1198); Pacing Problem (1439); Under-Sensing (1661)
Health Effect - Clinical Code Chest Pain (1776)
Date of Event 12/10/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
It was reported that the patient experienced chest pain.An interrogation report indicated that the right atrial (ra) lead position check failed and the ra lead exhibited undersensing resulting in false termination of atrial tachycardia/atrial fibrillation (at/af) detection, inappropriate pacing and misclassification of ventricular tachycardia non-sustained and supra ventricular tachycardia (svt) episodes.In addition, the inappropriate ra pacing lead to a blanked ventricular event that caused a functional loss of right ventricular (rv) lead capture of ventricular paced event.The ra and the rv lead remain in use.No further patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAPSUREFIX LEAD
Common Device Name
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
MPRI
road 149 km 56.3
villalba PR 00766
Manufacturer (Section G)
MPRI
road 149 km 56.3
villalba PR 00766
Manufacturer Contact
gerard torenvliet
8200 coral sea st ne
mounds view, MN 55112-4391
7635269277
MDR Report Key21897677
Report Number2649622-2025-11026
Device Sequence Number11623729
Product Code DTB
UDI-Device Identifier00613994776853
UDI-Public00613994776853
Combination Product (Y/N)N
Initial Reporter StateLA
Initial Reporter CountryUS
PMA/510(K) Number
P930039
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2012
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 04/24/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date06/09/2013
Device Model Number5568-53
Device Catalogue Number5568-53
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/03/2025
Initial Report FDA Received Date04/24/2025
Date Device Manufactured06/29/2011
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
5076-58 LEAD, W1DR01 IPG
Patient Age77 YR
Patient SexMale
Patient Weight113 KG
-
-