|
This supplemental report is being submitted to provide the results of the legal manufacturer's final investigation.Updated fields: d8, d9, h2, h3, h6, h11 the device was returned to olympus for inspection and the reported failure was confirmed.In addition, the following reportable malfunctions were identified during the device evaluation: images are not displayed, damage to the video connector receptacle, the lamp does not light up, two lamps are damaged.Based on the results of the investigation, it was likely that the most probable cause was traced to a component failure.However, a definitive root cause could not be identified.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|