| Catalog Number |
VTB042025 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
Device Embedded In Tissue or Plaque (3165)
|
| Date of Event |
04/22/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
In this event it is reported a profile vortex blue 04/20 25mm broke during use.The broken part was not retrieved and was incorporated into the fill.No injury.
|
| |
|
Additional Manufacturer Narrative
|
|
As a result of this malfunction, the potential for surgical intervention exists to preclude permanent damage to a body structure or permanent impairment of a body function as evidenced by previous reported events with similar files.This event, therefore, is reportable per 21cfr part 803.The device is available for evaluation, though has not been returned as of this report.Evaluation results will be submitted as they become available.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation completed.Customer returned 70x sealed files.Per ansi/aq z1.4-2003 inspector's rule (using single sampling plan 'normal' at inspection level s2 with aql 1.5), a sample lot of 8 were checked.Checked under microscope and found no manufacturing marks or defects.Files were measured using opt comp-007 (calibration date: (b)(6) 2025) and passed all attributes checked (wire size, d3, d6, and d11), see attached inspection form.Engineering reviewed and advised as no further testing required as sealed returned product meets specifications.Nothing unusual to report was found during dhr review.A query indicates no additional complaints have been received for this lot number.
|
| |
|