| Catalog Number |
VTB042525 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
Device Embedded In Tissue or Plaque (3165)
|
| Date of Event |
04/22/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
As a result of this malfunction, the potential for surgical intervention exists to preclude permanent damage to a body structure or permanent impairment of a body function as evidenced by previous reported events with similar files.This event, therefore, is reportable per 21cfr part 803.The device is available for evaluation, though has not been returned as of this report.Evaluation results will be submitted as they become available.
|
| |
|
Event or Problem Description
|
|
In this event it is reported that profile vortex blue 04/25 25mm broke during use.The broken part was not retrieved and was incorporated into the fill.No injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Customer returned 120x sealed files 45 of lot 411887 and 75 of lot 422397.Per ansi/aq z1.4-2003 inspector's rule (using single sampling plan 'normal' at inspection level s2 with aql 1.5), a sample lot of 8 were checked from each lot.Checked under microscope and found no manufacturing marks or defects.Files were measured using opt comp-007 (calibration date 2/24/2025) and passed all attributes checked (wire size, d3, d6, and d11), see attached inspection form.Also lab performed testing and all files tested met specifications.See attached lab testing documents.Returned product meets specifications.Engineering reviewed and advised no further testing needed at this time.Nothing unusual to report was found during dhr review.A query indicates no additional complaints have been received for these lot numbers.
|
| |
|
Search Alerts/Recalls
|