• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER DENTAL IMPL TAPERED SP 4.8MM 8MM OCTAGON; DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER DENTAL IMPL TAPERED SP 4.8MM 8MM OCTAGON; DENTAL IMPLANT Back to Search Results
Catalog Number SPWB8
Medical Device Problem Code Patient-Device Incompatibility (2682)
Health Effect - Clinical Codes Hypersensitivity/Allergic reaction (1907); Pain (1994)
Date of Event 03/10/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Zimvie complaint number (b)(6).G4: premarket identification pma/510(k) #: k011245|k002188.
 
Event or Problem Description
The doctor reports that the dental implant placed in position number #46 failed due to an allergic reaction three days after its placement in the mouth.The doctor also reports unbearable pain radiating from the jaw to the ear.
 
Event or Problem Description
No further event information is available at the time of this report.
 
Additional Manufacturer Narrative
Zimvie complaint number (b)(4) the following fields have been updated: b4: date of this report.B5: describe event or problem.G3: date received by manufacturer.G6: type of report.H1: type of reportable event.H2: follow up type.H3: device evaluated by manufacturer.H6: adverse event problem.H10: additional narrative.Zimvie received one (1) spwb8, (impl tapered sp 4.8mm 8mm octagon) for evaluation.Visual evaluation was performed, the implant had signs of use with debris on its internal and external threads.There was no damage identified or signs of malfunction that would have contributed to the event.Dhr review was completed for the subject lot number 2120019059.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Complaint history review was performed for the reported lot number 2120019059 for similar events and no other complaint was identified.Review completed utilizing keywords: dental : medical : allergic reaction and dental : medical : pain.Based on the investigation and risk management file review, the most likely root cause determined from the investigation was patient factors.Therefore, based on the available information, a device malfunction did not occur.No damage to the implant was identified.The reported event was non-verifiable.No further investigation or immediate capa / hhe/d escalation is required, as the complaint investigation did not confirm the products were nonconforming at the time of distribution, and no new failure mode, harm, or hazardous situation was identified through the investigation performed.At this time, the complaint investigation has been completed and the record will be closed.If additional information is received, the record will be re-opened for further evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPL TAPERED SP 4.8MM 8MM OCTAGON
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
ZIMMER DENTAL
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
ZIMMER DENTAL
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
carlos gordian-arroyo
4555 riverside drive
palm beach gardens, FL 33410
5619713230
MDR Report Key21899755
Report Number0002023141-2025-01075
Device Sequence Number4085224
Product Code DZE
UDI-Device Identifier00889024017597
UDI-Public(01)00889024017597(17)290920(10)2120019059
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K082639
Exemption Number5645646
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date (Section B) 08/05/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Catalogue NumberSPWB8
Device Lot Number2120019059
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/31/2025
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date04/24/2025
Supplement Report FDA Received Date08/05/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/21/2024
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age54 YR
Patient SexFemale
Patient Weight47 KG
Patient EthnicityNon Hispanic
-
-