Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.D4: udi: as a valid lot number for the device involved in the event was not provided, the full udi is currently not available.H3, h6 investigation summary : the product was not returned to medtech orthopaedics, however photos were provided for review.The photo investigation revealed that scrdriver shaft cann t8 w/colour-marking (03.226.004) had a broken tip.The observed condition of the device was consistent with a component failure that was caused by exposure to unintended forces like usage of excessive force in a prying motion while trialing.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.The overall complaint was confirmed as the observed condition of the scrdriver shaft cann t8 w/colour-marking would contribute to the complaint about the device issue.Based on the investigation findings, potential cause can be attributed to component failure and it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the complaint condition.As part of medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.The device lot number is unknown, therefore a device history review could not be performed.If the lot/serial number becomes available, the record will be re-assessed.
|
|
It was reported that during the procedure, while the doctor was screwing, the tip of the t screwdriver with color marking broke.A prompt solution was found to handle the situation by using another screwdriver, thus completing the surgery successfully, without affecting the patient, without discomfort to the specialist, and without alterations in surgical times.
|