Catalog Number 0168L16
Medical Device Problem Code Material Split, Cut or Torn (4008)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/14/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete, a supplemental report will be filed.Udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that the patient was presented for foley catheter exchange.Using new bard 16 fr coude catheter that had packaging intact, and was not expired, catheter was placed.It was then noted that urine was leaking out of a large tear/hole near the end of the catheter.This was promptly removed.
 
Additional Manufacturer Narrative
The reported event was confirmed-unknown cause.Unable to performed functional testing due to only photo attached and provided for evaluation.The potential root cause for this failure mode could be (example: user related/operator error/rough handling or expose to sharp object/mechanical force).A review of the device history record did not show any problems or conditions that would have contributed to the reported issue.The instructions for use were found adequate and states the following: warning: on catheter, do not use ointments or lubricants having a petrolatum base.They will damage the catheter and may cause balloon to burst.Warning: after use, this product may be a potential biohazard.Handle and dispose of in accordance with accepted medical practices and applicable local, state and federal laws and regulations.Visually inspect the product for any imperfections or surface deterioration prior to use.If package is opened or if any imperfection or surface deterioration is observed, do not use.Please consult product label and insert for any indications, contraindications, hazards, warnings, cautions and directions for use.Proper techniques for urinary catheter maintenance 1) secure the foley catheter.Use the statlock foley stabilization device if provided.2) maintain a closed drainage system by utilizing pre-connected, sealed catheter-tubing junctions.3) maintain unobstructed urine flow and keep the catheter and collection tube free from kinking.4) keep the collection bag below the level of the bladder or hips at all times.5) empty the collection bag regularly using a separate, clean collection container for each.Foley catheter removal 1) to deflate catheter balloon: gently insert a luer lock or slip tip syringe in the catheter valve.Never use more force than is required tomake the syringe ¿stick¿ in the valve 2) allow the pressure within the balloon to force the plunger back and fill the syringe with water.If you notice slow or no deflation, re-seat the syringe gently 3) use only gentle aspiration to encourage deflation if needed.Vigorous aspiration may collapse the inflation lumen, preventing balloon deflation 4) if the balloon will not deflate and if permitted by hospital protocol, the valve arm may be severed.If this fails, contact adequately trained professional for assistance, as directed by hospital protocol 5) should balloon rupture occur, care should be taken to assure that all balloon fragments have been removed from the patient.Correction- d, h udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that the patient was presented for foley catheter exchange.Using new bard 16 fr coude catheter that had packaging intact, and was not expired, catheter was placed.It was then noted that urine was leaking out of a large tear/hole near the end of the catheter.This was promptly removed.Per additional information received via email on 27jun2025, there was no impact and intervention to the patient due to the this of the device.
 
Event or Problem Description
It was reported that the patient was presented for foley catheter exchange.Using new bard 16 fr coude catheter that had packaging intact, and was not expired, catheter was placed.It was then noted that urine was leaking out of a large tear/hole near the end of the catheter.This was promptly removed.
 
Additional Manufacturer Narrative
The reported event was inconclusive because no sample was returned for evaluation.It is unknown whether the device had met relevant specifications.It was unknown whether the product had caused the reported failure.The potential root cause for this failure mode could be (example: user related/operator error/rough handling or expose to sharp object/mechanical force).A review of the device history record did not show any problems or conditions that would have contributed to the reported issue.The instructions for use were found adequate and state the following: warning: on catheter, do not use ointments or lubricants having a petrolatum base.They will damage the catheter and may cause balloon to burst.Warning: after use, this product may be a potential biohazard.Handle and dispose of in accordance with accepted medical practices and applicable local, state and federal laws and regulations.Visually inspect the product for any imperfections or surface deterioration prior to use.If package is opened or if any imperfection or surface deterioration is observed, do not use.Please consult product label and insert for any indications, contraindications, hazards, warnings, cautions and directions for use.Proper techniques for urinary catheter maintenance 1) secure the foley catheter.Use the statlock foley stabilization device if provided.2) maintain a closed drainage system by utilizing pre-connected, sealed catheter-tubing junctions.3) maintain unobstructed urine flow and keep the catheter and collection tube free from kinking.4) keep the collection bag below the level of the bladder or hips at all times.5) empty the collection bag regularly using a separate, clean collection container for each.Foley catheter removal 1) to deflate catheter balloon: gently insert a luer lock or slip tip syringe in the catheter valve.Never use more force than is required to make the syringe stick in the valve 2) allow the pressure within the balloon to force the plunger back and fill the syringe with water.If you notice slow or no deflation, re-seat the syringe gently 3) use only gentle aspiration to encourage deflation if needed.Vigorous aspiration may collapse the inflation lumen, preventing balloon deflation 4) if the balloon will not deflate and if permitted by hospital protocol, the valve arm may be severed.If this fails, contact adequately trained professional for assistance, as directed by hospital protocol 5) should balloon rupture occur, care should be taken to assure that all balloon fragments have been removed from the patient.Correction: d udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARDEX® LUBRICATH® FOLEY CATHETER COUDE TIP
Common Device Name
FOLEY CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
Manufacturer (Section G)
C.R. BARD INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington 30014
7707846100
MDR Report Key21900340
Report Number1018233-2025-02977
Device Sequence Number8729136
Product Code EZC
UDI-Device Identifier00801741017827
UDI-Public(01)00801741017827(11)241119(17)280531(10)NGJY1200
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K910846
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/25/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number0168L16
Device Lot NumberNGJY1200
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/07/2025
Supplement Date Received by Manufacturer06/27/2025
08/25/2025
Initial Report FDA Received Date04/24/2025
Supplement Report FDA Received Date06/30/2025
08/26/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/19/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-