| Model Number |
T510C33 |
| Medical Device Problem Codes |
Gradient Increase (1270); Insufficient Device Problem Information (3190); Patient Device Interaction Problem (4001); Device Stenosis (4066)
|
| Health Effect - Clinical Codes |
Endocarditis (1834); Tricuspid Valve Stenosis (2113); Insufficient Information (4580)
|
| Date of Event |
03/20/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that post implant of this 33mm hancock ii valve in the tricuspid position, it was explanted and replaced with a 31mm hancock ii valve in the tricuspid position.The reason for the replacement was not reported.No additional adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic received additional information that changed the event description and device relatedness of this previously reported event.There is no evidence to suggest the device caused or contributed to a serious injury.B5, b7, and h6 updated.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
Additional information was received that the reason for replacement was endocarditis.It was reported that a transthoracic echocardiogram (tte) performed prior to explant demonstrated normal biventricular function, severe stenosis of the tricuspid valve with mean gradient 12mmhg.It was stated that degeneration with large vegetations were noted on the valve.No additional adverse patient effects were reported.
|
| |
|