This supplemental report is being submitted to provide the results of the legal manufacturer's final investigation.Updated fields: d8, d9, h2, h3, h6, h11.The device was returned to olympus for inspection, and the reported failure was confirmed.In addition, the following reportable malfunction was identified during the device evaluation: disconnection of the filament.A root cause could not be identified.Based on the results of the investigation, it is likely that the following led to the malfunction: regarding the customer's allegation, the lamp did not light up, it was due to a component failure.The cause of the newly identified malfunction mentioned above was traced to component failure, too.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|