H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.Each event reported to bd is evaluated and investigated in accordance with our complaint investigation procedures.The investigation process includes, but is not limited to, evaluation of the event details provided by the complaint facility, a review of complaint history, manufacturing records and risk document where applicable, and an evaluation of the subject device when available to identify potential contributing factors.The device has not been returned to the manufacturer for evaluation.
|
|
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The complaint of a split in the molded joint is confirmed; however, the exact cause is unknown.One photograph of a 5 fr dual lumen powerpicc were returned for evaluation.An initial visual observation of the photograph showed a split in the molded joint just proximal the wings.The printing on the molded joint was observed to be faded.Use residue was observed on the sample.Slight damage was observed on the wings and extension leg tubing in the returned photograph.There were no distinguishing features in the returned photograph of the device which could aid in identification of a specific root cause.This complaint will be recorded for future trending and monitoring purposes.
|