MAUDE Adverse Event Report: ESTABLISHMENT LABS S.A. MOTIVA ERGONOMIX ROUND SILKSURFACE; MOTIVA IMPLANTS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ESTABLISHMENT LABS S.A. MOTIVA ERGONOMIX ROUND SILKSURFACE; MOTIVA IMPLANTS Back to Search Results
Catalog Number ERSF-355
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Ptosis (2620)
Date of Event 06/29/2024
Type of Reportable Event Serious Injury
Event or Problem Description
Korea - it was reported that a patient operated on in (b)(6) 2019, presented to the doctor in (b)(6) 2024 with bottoming out.No patient demographics, such as age, weight, or ethnicity, were provided by the reporter, and no adverse outcomes for the patient have been noted.Efforts to gather additional information are ongoing, and the investigation remains in progress.
 
Additional Manufacturer Narrative
Reported condition is a common and known reason for complaint, and it is clearly characterized in the product dfu included with the implant, as stated above.The instructions for use in the directions for use were reviewed to determine if there are indications that ensure the prevention and good handling of the product.For this event, it is considered that the information is suitable on the section of surgical precautions as follows: "the inferior displacement of a breast implant, increasing the distance between the nipple-areolar complex and the inframammary fold, after breast implant surgery.Risk factors reported in the literature are, but not limited to the quality of the preexisting breast tissue (thin subcutaneous tissue, defective dermal elements, breast tuberosity), the breast implant selection (larger implants), the imf dissection and the placement of the implant during surgery (submuscular and subglandular planes).The clinical symptoms resulting from a bottoming out implant are asymmetry, upward-pointing nipples, sagging breast, palpable implant, among others.Prevention of this complication is based on the anticipation of the possible causes, for example: a careful and individual assessment of the mammary soft-tissues, a careful implant selection, preparation with a technique that can minimize the risk, and to provide adequate breast support after the surgery.The treatments may vary depending on the severity of the complication and go from a simple sub mammary fixation to the use of additional supporting materials." also, a complete review of the dhr's was carried out, no deviation was found in the process, so it is established that there is no evidence of a relationship between the event and the manufacturing process.
 
Additional Manufacturer Narrative
1.Overview the quality department (pms) received a complaint involving a motiva® device.2.General conclusion device involvement: a causal relationship between the reported events and the device has been established.Event characterization: the event was classified as bottoming out.3.Clinical confirmation upon thorough evaluation of the submitted clinical documentation and supporting evidence, the reported bottoming out was confirmed by clinical affairs according to the clinical evidence provided.Root cause analysis this event is a well-documented complication inherent to breast implant surgery.Based on the analysis of the available data, including clinical findings and product traceability records, there is no evidence indicating a causal link between this specific case and any product-related factors, including raw materials or manufacturing processes.The etiology of the alleged event is recognized as multifactorial, given the absence of manufacturing deviations or anomalies, and in line with current clinical literature, the event is considered not attributable to the manufacturing process and/or the product itself.4.Device history record (dhr) review a comprehensive review of the device history record (dhr) for the affected lot was completed.No deviations, non-conformances, or anomalies were identified during manufacturing.All raw materials and process parameters conformed to specified quality requirements.5.Trend and signal monitoring the event has been evaluated as part of ongoing trend analysis activities within the pms program.Per the current complaint data report: no adverse or unexpected trends related to device rupture have been identified.No further corrective or preventive actions are required at this time.Establishment labs will continue to monitor for similar events as part of routine pms surveillance.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MOTIVA ERGONOMIX ROUND SILKSURFACE
Common Device Name
MOTIVA IMPLANTS
Manufacturer (Section D)
ESTABLISHMENT LABS S.A.
building b-25, coyol free zone
alajuela,alajuela
Manufacturer (Section G)
ESTABLISHMENT LABS S.A.
building b-25, coyol free zone
alajuela,alajuela
Manufacturer Contact
susana hidalgo
building b-25, coyol free zone
alajuela,alajuela 
8003554164
MDR Report Key21903984
Report Number3012883202-2025-00605
Device Sequence Number4085356
Product Code FTR
UDI-Device Identifier07445161010766
UDI-Public(01)07445161010766(11)190221(17)240221
Combination Product (Y/N)N
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/08/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date02/21/2024
Device Catalogue NumberERSF-355
Device Lot Number19020287
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/04/2025
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date04/24/2025
Supplement Report FDA Received Date08/08/2025
Date Device Manufactured02/21/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-