| Model Number |
PCF-PH190I |
| Medical Device Problem Code |
Device Reprocessing Problem (1091)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/31/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Based on the results of the investigation, olympus confirmed foreign material came out of the device, but the type of the material cannot be identified however, it is likely the following led to the malfunction: the foreign material was possibly not removed completely if the reprocessing steps were not conducted properly.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|
Event or Problem Description
|
|
It was observed that during the device evaluation, the subject device exhibited foreign material in air/water tube and cylinder.There was no patient involvement.
|
| |
|
Search Alerts/Recalls
|