This is report 1 of 4.Correction: h3.Yes.H6.Investigation findings: 3251.Investigation conclusion: 4307.Device evaluation: visual inspection confirms the event.The screw fractured at the cam feature.Cam failure is typically caused by fatigue fracture through dynamic loading of the screw in tension under high loads.Review of device history records showed no manufacturing or processing related irregularities.Labeling review: warnings/cautions/precautions: risks identified with the use of these devices, which may require additional surgery, include device component failure, loss of fixation/stabilization, non-union, vertebral fracture, neurological injury, vascular or visceral injury.Possible adverse effects: initial or delayed loosening, disassembly, bending, dislocation, and/or breakage of device components.Postoperative management: the surgeon should instruct the patient regarding amount and time frame after surgery of any weight bearing activity.The increased risk of bending, dislocation, and/or breakage of the implant devices, as well as an undesired surgical result are consequences of any type of early or excessive weight bearing, vibratory motion, fall, jolts or other movements preventing proper healing and/or fusion development.Implant devices should be revised or removed if bent, dislocated, or broken.Immobilization should be considered in order to prevent bending, dislocation, or breakage of the implant device in the case of delayed, mal-union, or non-union of bone.Immobiliz should continue until a complete bone fusion mass has developed and been confirmed.
|