| Catalog Number |
07P57-30 |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/27/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer observed falsely elevated calcium generated from alinity c processing module.The customer compared results with other analyzers, including the architect platform and provided the following data: customer reference range for adjusted calcium is 2.2-2.6mmol/l.Customer reference range for calcium is 1.75-3mmol/l.Sample id (b)(6).Results on alinity were 3.04 (on sn: (b)(6), 4.78 (on sn:(b)(6), & 4.30 mmol/l (on sn: (b)(6).Result on architect were 2.12 & 2.49 mmol/l.Per the customer discrepant results were not reported out to the medical provider.There was no reported impact to patient management.
|
| |
|
Additional Manufacturer Narrative
|
|
An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.
|
| |
|
Additional Manufacturer Narrative
|
|
The complaint investigation included a review of data and information provided by the customer, a review of tracking and trending, labeling review, and device history record review.Return testing was not completed as returns were not available.Data and information provided by the customer were reviewed and support the complaint issue without indication for any additional issue.Review of tracking and trending for the alinity c calcium assay did not identify an increase in complaint activity related to the complaint issue.Additionally, an increase in complaint activity was not identified for reagent complaint lot.Device history record review did not identify any nonconformances or deviations with complaint lot and the complaint issue.Labeling was reviewed and sufficiently addresses the customer's issue.Based on our investigation, no systemic issue or deficiency with the alinity c calcium reagent lot 25020un24 was identified.
|
| |
|
Event or Problem Description
|
|
The customer observed falsely elevated calcium generated from alinity c processing module.The customer compared results with other analyzers, including the architect platform and provided the following data: customer reference range for adjusted calcium is 2.2-2.6mmol/l.Customer reference range for calcium is 1.75-3mmol/l.Sample id (b)(6).Results on alinity were 3.04 (on sn: (b)(6), 4.78 (on sn: (b)(6), & 4.30 mmol/l (on sn: (b)(6).Result on architect were 2.12 & 2.49 mmol/l.Per the customer discrepant results were not reported out to the medical provider.There was no reported impact to patient management.
|
| |
|
Search Alerts/Recalls
|