• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS, INC. EVERFLO; GENERATOR, OXYGEN, PORTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. EVERFLO; GENERATOR, OXYGEN, PORTABLE Back to Search Results
Model Number 1020011
Medical Device Problem Code Smoking (1585)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/11/2025
Type of Reportable Event Malfunction
Event or Problem Description
The manufacturer was contacted on everflo device that came back from repairs a week ago, but it smells like smoke when it is in operation.No medical intervention was specified as required by the patient.The device has not yet been returned to the manufacturer for evaluation.The manufacturer's investigation is ongoing.A supplemental follow-up report will be submitted when the manufacturer's investigation is complete.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVERFLO
Common Device Name
GENERATOR, OXYGEN, PORTABLE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer Contact
melissa rosko
6501 living place
pittsburgh, PA 
4125423300
MDR Report Key21908266
Report Number2518422-2025-104526
Device Sequence Number15658825
Product Code CAW
UDI-Device Identifier00606959032590
UDI-Public(01)00606959032590
Combination Product (Y/N)N
Initial Reporter CountryIT
PMA/510(K) Number
K061261
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 04/25/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number1020011
Device Catalogue Number1020011
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/11/2025
Initial Report FDA Received Date04/25/2025
Date Device Manufactured06/25/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-