MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION LIKORALL¿ OVERHEAD LIFT; LIFT, PATIENT, NON-AC-POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION LIKORALL¿ OVERHEAD LIFT; LIFT, PATIENT, NON-AC-POWERED Back to Search Results
Model Number 3124050
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/08/2025
Type of Reportable Event Malfunction
Event or Problem Description
A company representative - service personnel contacted technical service to report that the likorall 250 s, natural, irc emergency stop does not work if it is pushed.This occurred before use.There was no patient/user injury, medical intervention, symptom, or adverse event reported.There were no other devices reported to be involved.The customer did not communicate this event to another baxter department or authority.The device was requested for service/inspection by baxter.
 
Additional Manufacturer Narrative
Upon inspection, the baxter technician found the emergency stop button needed to be replaced.Likorall overhead lift is intended for use in, health care, intensive care and rehabilitation.Likorall overhead lift is designed for fixed installation and free-standing lift systems.All common lifts and transfers can be performed using likorall overhead lift, for instance between bed/wheelchair, to/from floor, toilet visits, gait training, and together with stretchers.Likorall r2r (room to room) overhead lift enables the patient to be moved between two rail systems in separate rooms.Likorall overhead lift with the es designation is prepared for operation with the wireless handcontrol remote (ir) and in addition, a transfer motor can be connected for motor driven movement along the rail.Likorall s, irc overhead lift is prepared for continuous charging through the railsystem.Although there was no adverse event associated with this incident, if the emergency stop button were to be faulty during a patient transfer, it could lead to serious injury or death.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIKORALL¿ OVERHEAD LIFT
Common Device Name
LIFT, PATIENT, NON-AC-POWERED
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
LIKO AB
nedre vagen 100
lulea 975 9 2
SW   975 92
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key21908275
Report Number8030916-2025-00066
Device Sequence Number4008952
Product Code FSA
UDI-Device Identifier00887761979567
UDI-Public(01)00887761979567
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 04/25/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number3124050
Device Catalogue Number3124050
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/09/2025
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/08/2025
Initial Report FDA Received Date04/25/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-