MAUDE Adverse Event Report: ARROW INTERNATIONAL LLC ARROW FIBEROPTIX ULTRA 8 IAB: 8FR 30CC; SYSTEM, BALLOON, INTRA-AORTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARROW INTERNATIONAL LLC ARROW FIBEROPTIX ULTRA 8 IAB: 8FR 30CC; SYSTEM, BALLOON, INTRA-AORTIC Back to Search Results
Catalog Number IAB-06830-U
Medical Device Problem Code Inflation Problem (1310)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 06/14/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported "balloon was causing high pressure alarms potentially because it was not unraveling all the way.Balloon was removed".No patient injury or consequence reported.At the time of this report, the customer has not returned our requests for additional information.If further information is received, the complaint file will be updated.
 
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
It was reported "balloon was causing high pressure alarms potentially because it was not unraveling all the way.Balloon was removed".No patient injury or consequence reported.At the time of this report, the customer has not returned our requests for additional information.If further information is received, the complaint file will be updated.
 
Additional Manufacturer Narrative
(b)(4).The serial number on the returned sample is (b)(6).Returned for investigation was a 30cc 7.5fr ultraflex intra-aortic balloon catheter (iabc) without the original packaging.The sample was returned in the ups shipping box and was in a sealed bio-hazard bag.Upon return, the teflon sheath was noted on the iabc.The one-way valve was tethered to the short driveline tubing.The supplied 30cc inflation driveline tubing was connected to the iabc short driveline tubing.The supplied short arterial pressure tubing was noted connected to the iabc luer.The bladder was fully unwrapped.A bend to the iabc central lumen was noted at approximately 59cm from the iabc distal tip.Dried blood was noted on the exterior surfaces of the returned iabc.No blood was noted within the helium pathway.The bladder thickness was measured at six points with measurements ranging from 0.0061in-0.0065in and was within specification of process document.The one-way valve was tested and passed.A vacuum was pulled on the one-way valve, and it held for at least 1 minute and then 30 seconds five separate times according to quality system document.The iabc central lumen was aspirated and flushed using a 60cc lab-inventory syringe.No abnormalities or debris were noted.The iabc was connected to an iabp with the returned 30cc inflation driveline tubing.The iabc was inserted into the "t" tube and 75mmhg backpressure was applied.The iabc was pumped for a minimum of 15 minutes.There were no alarms triggered.The bladder inflated and deflated completely with each beat.The iabc was leak tested in accordance with testing methods from manufacturing procedure.No leaks were detected.Full inflation was achieved.The device passed the leak test.A lab inventory 0.025in guidewire was back loaded through the iabc distal tip.Resistance was noted at approximately 58.7cm from the iabc distal tip, which is the location of the previously noted bend.The guidewire was able to advance through the central lumen.No blood or debris was noted.A de-vice history record (dhr) review was conducted for the lot number with no relevant findings.The device passed all manufacturing specifications prior to release.The reported complaint for iab "not unraveling all the way" is not confirmed.The returned iabc bladder was fully intact with no abnormalities noted.During the investigation, the iabc fully inflated, and no leaks/alarms were noted.The returned device passed visual and functional test specifications.Based on a review of the device history record (dhr), the product met specification upon release.The root cause of the complaint is undetermined.No further action required at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARROW FIBEROPTIX ULTRA 8 IAB: 8FR 30CC
Common Device Name
SYSTEM, BALLOON, INTRA-AORTIC
Manufacturer (Section D)
ARROW INTERNATIONAL LLC
morrisville NC
Manufacturer (Section G)
ARROW INTERNATIONAL INC.
16 elizabeth drive
chelmsford MA 01824
Manufacturer Contact
sachin sony
3015 carrington mill blvd
morrisville, NC 27560
MDR Report Key21908986
Report Number3010532612-2025-00330
Device Sequence Number12967098
Product Code DSP
UDI-Device Identifier00801902034724
UDI-Public(01)00801902034724(11)230818(17)250731(10)18F23H0059(21)
Combination Product (Y/N)N
Initial Reporter StateOR
Initial Reporter CountryUS
PMA/510(K) Number
K021462
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/01/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Expiration Date07/31/2025
Device Catalogue NumberIAB-06830-U
Device Lot Number18F23H0059
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/01/2025
Supplement Date Received by Manufacturer04/30/2025
Initial Report FDA Received Date04/25/2025
Supplement Report FDA Received Date05/08/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/18/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-