• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BARD® URINARY DRAINAGE BAG; DRAIN BAG

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BARD® URINARY DRAINAGE BAG; DRAIN BAG Back to Search Results
Catalog Number 153509
Medical Device Problem Code Restricted Flow rate (1248)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/09/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.Udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that the urine flows back to the catheter and the urine did not drain properly (eod on (b)(6) 2025, eod on (b)(6) 2025).
 
Event or Problem Description
It was reported that the urine flows back to the catheter and the urine did not drain properly (eod-(b)(6) 2025, eod-(b)(6) 2025).
 
Additional Manufacturer Narrative
The reported event was inconclusive because this investigation did not result in any additional findings, and no sample was available for evaluation.No actions can be taken at this time since a root cause was not identified.A review of the device history record did not show any problems or conditions that would have contributed to the reported issue.The instructions for use were found adequate and state the following: visually inspect the product for any imperfections or surface deterioration prior to use.If package is opened or if any imperfection or surface deterioration is observed, do not use.1.Remove protective cap from drainage tube catheter adapter and connect drainage tube to catheter.2.Position hanger on bedside rail, using string or hook.3.Use sheeting clip to secure drainage tube to sheet.Important: hang drainage tube in a straight fashion from bedside to drainage bag.Ensure that the drainage bag is placed near the foot of the bed.4.If using a urine meter, it may be emptied in two ways: a.To empty into the bag, grasp the bottom of the meter and lift up.To ensure that the meter empties completely, lifting again is recommended.B.To empty urine meter into receptacle, twist green portion of drain valve to the left; to close, twist green position of the drain valve to the right.5.To empty bag: a.Remove outlet tube from housing; gently squeeze connector arms and pull tube from housing.B.Release clamp and empty bag.C.After emptying, reclamp outlet tube and slide connector into housing until connector arms engage.Note: if specimen is required, see directions for using urine sample port.6.Periodic observations of this system should be made to ensure that urine is flowing freely.If a standing column of urine is observed, check for correct positioning of bag and then for a physical obstruction.If correct positioning or removal of physical obstruction does not allow free flow, the bag may have to be changed.7.If bag is not positioned correctly, urine may bypass the meter and go directly to the bag.Reposition bag as necessary.Directions for using bard ez-lok sampling port: bard ez-lok sampling port accepts a luer-lock or slip tip syringe.1.Kink drainage tubing a minimum of 3 inches below the sampling port until urine is visible under the access site.2.Swab surface of site with antiseptic wipe.3.Using aseptic technique, position the syringe in the center of the sampling port.The syringe should be held perpendicular to the surface of the sampling port (at approximately 80-100 degree angle).Press the syringe firmly and twist gently to lock the syringe onto the sampling port.Note: improper penetration technique could cause formation of a drop of urine on the surface of the sampling port.Periodic observation of the sampling port is recommended.4.Aspirate desired volume of urine.5.Unkink tubing and send specimen to laboratory.Correction: d.Udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD® URINARY DRAINAGE BAG
Common Device Name
DRAIN BAG
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
Manufacturer (Section G)
C.R. BARD INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington 30014
7707846100
MDR Report Key21909290
Report Number1018233-2025-03003
Device Sequence Number4221807
Product Code KNX
UDI-Device Identifier00801741029608
UDI-Public(01)00801741029608(11)250115(17)291031(10)NGJY1752
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Consumer,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 07/23/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Catalogue Number153509
Device Lot NumberNGJY1752
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/09/2025
Supplement Date Received by Manufacturer07/23/2025
Initial Report FDA Received Date04/25/2025
Supplement Report FDA Received Date07/24/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured01/15/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-