• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COLOPLAST A/S TOROSA SALINE-FILLED TESTICULAR PROSTHESIS; PROSTHESIS, TESTICULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COLOPLAST A/S TOROSA SALINE-FILLED TESTICULAR PROSTHESIS; PROSTHESIS, TESTICULAR Back to Search Results
Catalog Number 450-1327
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/28/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The lot number was reviewed for complaint trend, non-conforming report and capa.Devices met specification prior to release.No trends were noted for complaints and there were no non-conforming reports or capas that were confirmed to be associated.The device was not returned for evaluation.Without the benefit of analyzing the device, quality cannot comment on the condition of the device.Should additional information prompt us to alter or supplement any information or conclusions contained in this report, a follow-up report will be submitted.
 
Event or Problem Description
According to the available information, prior to patient use / patient contact, the implant leaked.
 
Additional Manufacturer Narrative
Testicular implant was received for evaluation.A separation was noted in the shell of the testicular.This was a site of leakage.The separation appeared to have a central groove, indicating contact with sharp instrumentation such as a needle.Because this component was released according to manufacturing and quality control procedures, it was concluded that the observed instrument separation in the testicular shell occurred subsequent to the device packaging being opened.The testicular device was returned with saline inside.Based on the evaluation and information received, it was concluded that the separation most likely occurred inadvertently during the filling of the device prior to implant.A review of the device history record confirmed the devices from this lot met all specifications prior to release.A review of the complaint history database, nonconformances and capas revealed no trends for this lot.
 
Event or Problem Description
According to the available information, prior to patient use / patient contact, the implant leaked.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TOROSA SALINE-FILLED TESTICULAR PROSTHESIS
Common Device Name
PROSTHESIS, TESTICULAR
Manufacturer (Section D)
COLOPLAST A/S
1 holtedam humlebaek, dk 3050
humlebaek 3050
DA  3050
Manufacturer Contact
usbes brian schmidt
1601 west river road n
minneapolis, MN 55411
MDR Report Key21909739
Report Number2125050-2025-00698
Device Sequence Number4009006
Product Code FAF
UDI-Device Identifier05708932470834
UDI-Public05708932470834
Combination Product (Y/N)N
Initial Reporter CountryNZ
PMA/510(K) Number
P020003
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/03/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number450-1327
Device Lot Number9528751_5206302400
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date04/25/2025
Supplement Report FDA Received Date09/03/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/25/2024
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-