| Model Number |
SUTR ANCH,CRKSCRW |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/10/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation is in process.A follow-up report will be provided upon availability of additional information.
|
| |
|
Event or Problem Description
|
|
It was reported that during a rotator cuff suture surgery, after the anchor was inserted, it was found that the eylet of the anchor was defective and therefore the fiberwire became detached from the anchor.It was further reported that the anchor remained inside the patient.Per complaint reporter there was no harm for patient, operator or third party.The surgery was finished successfully with a new device with the same part number.It was not necessary to switch the surgical technique or do a second surgery.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: g3, h3, h6 based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause for the reported failure can be attributed to user error of the device due to excessive force being used during insertion of the implant.
|
| |
|