| Catalog Number |
9770700 |
| Medical Device Problem Code |
Poor Quality Image (1408)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/07/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Observation found during evaluation at bd service center: there are horizontal lines in the ultrasound image due to interference within the rfid antenna.This is due to an internal failure within the rfid antenna.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The date of event was not provided by the complainant/reporter, the date reflected in this report is the date bd became aware of the event.The manufacturer has received the sample and is pending evaluation.A supplemental will be submitted with evaluation results.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The device was returned to the service facility for evaluation.Observation found during evaluation: there are horizontal lines in the ultrasound image due to interference within the rfid antenna.This is due to an internal failure within the rfid antenna was confirmed.Root cause is an internal failure of the rfid antenna.
|
| |
|
Search Alerts/Recalls
|