| Model Number |
9390A-1001SP |
| Medical Device Problem Code |
Unable to Obtain Readings (1516)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Complainant alleged that during functional testing, the device was unable to obtain an ecg signal via electrode pads.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to zoll medical corporation for evaluation.The customer's report was observed during review of the device data logs.However, the device was put through extensive testing including on/off current, patient and circuit impedance testing, and shock testing without duplicating the report.An internal inspection found no discrepancies.The main board was replaced as a precaution.The board was scrapped after testing.The device was recertified and returned to the customer.Analysis of reports of this type has not identified an increase in trend.
|
| |
|
Search Alerts/Recalls
|