| Catalog Number |
1115000000X |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/01/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This record is a consolidation of records summarized as part of the fda voluntary malfunction summary reporting program.2 devices were functionally/visually inspected in the field.The devices were repaired and returned to use.1 device was not evaluated and no cause was determined, as the customer did not make the device accessible for testing.1 device is pending evaluation.There was no remedial action taken.This device is not labeled for single use.
|
| |
|
Event or Problem Description
|
|
This report summarizes 4 malfunction events, where it was reported the devices experienced fluid leaks onto floor.No adverse consequence or clinically relevant delay in treatment was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The device that was pending evaluation was evaluated in the field and the issue was confirmed.The device was repaired on site and returned to service.Section h codes have been updated to reflect this.
|
| |
|
Event or Problem Description
|
|
No new information.
|
| |
|
Search Alerts/Recalls
|