Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|
H3: product analysis ¿ as found condition: the pipeline flex shield embolization device and phenom 27 catheter were returned for analysis within a shipping box; and within a plastic bio-pouch.The pipeline flex shield device was returned outside the phenom 27 catheter.¿ damage location details: no bent or kink was found with pushwire.The distal hypotube and ptfe shrink tubing were found to be intact with no signs of elongation.The distal and proximal dps restraints were found to be intact.The dps sleeves were found intact with no signs of damage.No defects were found with the tip coil, distal marker, re-sheathing marker, arms or with the proximal bumper.The braid was returned already detached from the pusher; therefore, the proximal and distal ends could not be identified.One of the ends of pipeline flex shield braid appeared to be not fully opened due to damaged braid.The other end of the pipeline flex shield braid was found fully opened and damaged.In addition, the middle section was found to be fully opened and no damage.No flash or voids molded were observed in the hub.The catheter body was found to be flattened from 2.8cm to 4.2cm from distal end.The catheter tip, marker band were examined; and no damages were found.No other anomalies were observed.¿ testing/analysis: the total and usable lengths of the catheter were measured to be within specifications.The catheter was flushed with water and water exited out from the distal tip.The catheter was then tested by running an in-house 0.0265¿ mandrel through catheter tip and hub.The mandrel successfully passed through the catheter hub and tip with no issue.¿ conclusion: based on the analysis findings, the pipeline flex shield could not be confirmed to have failure /incomplete open at middle section as the middle section was found to be fully opened and no damage.However, one of the ends of pipeline flex shield braid appeared to be not fully opened due to damaged braid.Possible causes of failure/incomplete to open include vessel tortuosity and damage braid.In addition, based on the analysis findings, the pipeline flex shield and phenom 27 were confirmed to have resistance during retrieval as the pipeline flex shield braid and phenom 27 were found to be damaged.Possible causes include incompatible catheter, damaged catheter, burrs, bumps, kinks or other damage on ped or pushwire, frayed ends on braid, patient vessel tortuosity, user does not maintain continuous flush, and users pulls back on/torques wire while advancing ped in microcatheter.It is likely these damages occurred when the customer attempted to advance and retrieve the pipeline flex shield through the phenom 27 catheter against resistance medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|