• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER MEDICAL-KALAMAZOO EYE SURGERY STRETCHER; STRETCHER, WHEELED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER MEDICAL-KALAMAZOO EYE SURGERY STRETCHER; STRETCHER, WHEELED Back to Search Results
Catalog Number 1089000000
Medical Device Problem Code Device Fell (4014)
Health Effect - Clinical Codes Fall (1848); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2025
Type of Reportable Event Malfunction
Event or Problem Description
This report summarizes 4 malfunction events, where it was reported the devices experienced head piece no longer supports weight or siderail unexpected drop/collapse.There was 1 event with patient involvement with a fall; no adverse consequence or clinically relevant delay in treatment was reported.
 
Additional Manufacturer Narrative
This record is a consolidation of records summarized as part of the fda voluntary malfunction summary reporting program.3 devices were functionally/visually inspected in the field.The devices were repaired and returned to use.1 device was not evaluated, as a probable cause for the issue was identified during communication between the customer and stryker and no further assistance was requested by the customer.There was no remedial action taken.This device is not labeled for single use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EYE SURGERY STRETCHER
Common Device Name
STRETCHER, WHEELED
Manufacturer (Section D)
STRYKER MEDICAL-KALAMAZOO
3800 east centre avenue
portage MI 49002
Manufacturer (Section G)
STRYKER MEDICAL-KALAMAZOO
3800 east centre avenue
portage MI 49002
Manufacturer Contact
melissa simon
3800 east centre avenue
portage, MI 49002
2693292100
MDR Report Key21912280
Report Number0001831750-2025-00445
Device Sequence Number13022354
Product Code FPO
UDI-Device Identifier07613327278170
UDI-Public07613327278170
Combination Product (Y/N)N
Initial Reporter CountryUS
Exemption NumberVMSR
Number of Events Summarized4
Summary Report (Y/N)Y
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 04/25/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue Number1089000000
Was Device Available for Evaluation? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/01/2025
Initial Report FDA Received Date04/25/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-