• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. SVCE REP, DYONICSPOWERMINI,HAND CONTROLS; SAW, POWERED, AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. SVCE REP, DYONICSPOWERMINI,HAND CONTROLS; SAW, POWERED, AND ACCESSORIES Back to Search Results
Catalog Number 72201500S
Medical Device Problem Code Physical Resistance/Sticking (4012)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H11: internal complaint reference: (b)(4).H3, h6: this complaint was opened by smith+nephew to document a product problem associated with a smith+nephew device.The reported problem relates to known inherent device and/or procedural risks that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are conservatively submitting this report in accordance with applicable regulations.If additional investigative findings or information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
 
Event or Problem Description
Upon investigation it was determined that the dyonics powermini shaver's button was stuck causing the shaver to run continuously.There was no patient involvement.
 
Additional Manufacturer Narrative
H3, h6: the reported device was received for evaluation.A visual inspection revealed no physical damage.A functional evaluation was performed on the returned device and found a stuck button causing device to run continuously.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A complaint history review found no similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.The root cause has been associated with a component failure.Factors that could have contributed to the failure include a buildup of corrosion/debris around the spring from cleaning chemical fluid ingression over time.Based on this investigation, the need for corrective action is not indicated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SVCE REP, DYONICSPOWERMINI,HAND CONTROLS
Common Device Name
SAW, POWERED, AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107
Manufacturer (Section G)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107
Manufacturer Contact
holly topping
1450 brooks rd.
memphis, TN 38116
5123913905
MDR Report Key21914047
Report Number1643264-2025-00138
Device Sequence Number7109315
Product Code HAB
UDI-Device Identifier03596010654823
UDI-Public3596010654823
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/29/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number72201500S
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/03/2025
Supplement Date Received by Manufacturer05/28/2025
Initial Report FDA Received Date04/25/2025
Supplement Report FDA Received Date05/29/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-