|
Zimvie complaint number (b)(4).A1: patient identifier unknown / not provided.D10.Concomitant medical products boet413 , 3i t3 ex hex tapered implant 4 x 13mm lot.2016041578 & boet510, 3i t3 ex hex tapered implant 5 x 10mm lot 2016051560.E1: reporter email address unknown / not provided.H6: event problem codes ¿ patient code: 1932 inflammation.A summary investigation has been completed for peri-implantitis events recognizing that a definitive root cause cannot be identified due to a wide range of external factors (non-design or manufacturing related), including medical conditions (e.G., diabetes, poor bone quality, etc.) / patient habits (e.G., smoking) and surgical technique.Previously completed investigations for these events have not identified any signals indicating potential non-conformances affecting the manufacturing and sterilization processes.Furthermore, the probability of a manufacturing or design defect that might lead to peri-implantitis occurring and escaping the available detections has been assessed and found remote and almost nonexistent.Should additional information be received which indicates that the device may have caused or contributed to the event, an additional report will be submitted.
|