ZIMMER BIOMET, INC. COMPR RVS GLENO 2-PRNG INS/IMP; ORTHOPAEDIC INORGANIC IMPLANT INSERTER/EXTRACTOR, REUSABLE Back to Search Results
Catalog Number 110028879
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/09/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the pin at the tip of the device was found to be fractured/missing prior to use during a procedure.No patient impact was reported.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Additional Manufacturer Narrative
(b)(4).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated: b4, b5, g3, g6, h2, h3, h6, h11.Visual examination of the provided photos show a circle around the part number reported.Review of the device history record(s) identified no deviations or anomalies during manufacturing.A definitive root cause cannot be determined.The reported event is not confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information is available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COMPR RVS GLENO 2-PRNG INS/IMP
Common Device Name
ORTHOPAEDIC INORGANIC IMPLANT INSERTER/EXTRACTOR, REUSABLE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key21917627
Report Number0001825034-2025-01218
Device Sequence Number8907548
Product Code PHX
UDI-Device Identifier00880304731394
UDI-Public(01)00880304731394(11)180801(10)801370
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
K193373
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/22/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number110028879
Device Lot Number801370
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/09/2025
Supplement Date Received by Manufacturer08/14/2025
Initial Report FDA Received Date04/28/2025
Supplement Report FDA Received Date08/22/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/01/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexMale
-
-