• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. 30MM CENTRAL SCREW, MODULAR; SHOULDR PROSTH, REVERSE CONFIG

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. 30MM CENTRAL SCREW, MODULAR; SHOULDR PROSTH, REVERSE CONFIG Back to Search Results
Model Number 30MM CENTRAL SCREW, MODULAR
Medical Device Problem Codes Positioning Failure (1158); Defective Device (2588)
Health Effect - Clinical Codes Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/28/2025
Type of Reportable Event Malfunction
Event or Problem Description
On 4/2/2025, it was reported by a sales representative via email that an ar-9560-24-2 arthrex univers revers baseplate and an ar-9561-30s arthrex univers revers central screw failed to properly mate after being pressed together while on the back table.When attempting implantation, the baseplate repeatedly disengaged from the screw and separated.This resulted in the inability to use the items as intended.This was discovered during a reverse total shoulder procedure on (b)(6) 2025.
 
Additional Manufacturer Narrative
Investigation is in process.A follow-up report will be provided upon availability of additional information.
 
Event or Problem Description
Additional information was received on 05/05/2025: the case was completed using a new ar-9560-24 arthrex univers revers baseplate and an ar-9561-30p univers revers central post instead of a screw.The case was delayed 5 minutes with no added anesthesia administered to the patient.
 
Additional Manufacturer Narrative
Additional information: b5, g3, h3, h6.Based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause for the reported failure can be attributed to misuse/mishandling of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
30MM CENTRAL SCREW, MODULAR
Common Device Name
SHOULDR PROSTH, REVERSE CONFIG
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
jared engle
8009337001
MDR Report Key21917638
Report Number1220246-2025-01769
Device Sequence Number18413600
Product Code PHX
UDI-Device Identifier00888867296572
UDI-Public(01)00888867296572(11)240726(17)290630(10)15293607
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
K173900
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 05/28/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number30MM CENTRAL SCREW, MODULAR
Device Catalogue NumberAR-9561-30S
Device Lot Number15293607
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/02/2025
Supplement Date Received by Manufacturer05/28/2025
Initial Report FDA Received Date04/28/2025
Supplement Report FDA Received Date05/28/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured07/26/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-