| Model Number |
6900PTFX |
| Medical Device Problem Code |
Device Stenosis (4066)
|
| Health Effect - Clinical Codes |
Dyspnea (1816); Heart Failure/Congestive Heart Failure (4446)
|
| Date of Event |
04/02/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
H11: additional manufacturer narrative: the subject device is not available for evaluation, as it remains implanted in the patient.The investigation is still in progress; therefore, a conclusion has yet to be established.A supplemental report will be submitted accordingly upon investigation completion.
|
| |
|
Event or Problem Description
|
|
It was reported that a 27mm 6900ptfx mitral valve in mitral position was disabled via valve in valve procedure after an implant duration of 9 years, 1 month due to stenosis.The patient presented with heart failure and shortness of breath.Tmvr was completed with a 26mm 9755rsl transcatheter valve and patient was in stable condition post-operative.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: additional manufacturer narrative: the device history record (dhr) review was completed, and this device passed all manufacturing and sterilization inspections prior to release for distribution.There were no issues identified that would have impacted this event.Engineering evaluation summary: the subject device will not be returned for evaluation as it remains implanted.Per doc-0249432, stenosis is a common manifestation of bioprosthetic valve and valved conduit dysfunction and has many potential root causes, including structural valve deterioration (svd), non-structural valve dysfunction (nsvd), endocarditis and/or thrombosis.Bioprosthetic device stenosis is typically related to patient and/or procedural factors and is not typically an indication of a device malfunction related to a manufacturing deficiency.There is no evidence to suggest a manufacturing defect caused or contributed to the reported event.The most likely root cause is patient factors, including diabetes mellitus and coronary artery disease.
|
| |
|
Search Alerts/Recalls
|