| Model Number |
8.0 MM FLIP CUTTER II |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/15/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation is in process.A follow-up report will be provided upon availability of additional information.
|
| |
|
Event or Problem Description
|
|
It was reported that during an anterior cruciate ligament surgery the tip of the flip cutter broke.No part of the device remained inside the patient.Per complaint reporter there was no harm for patient, operator or third party.The surgery was finished successfully with a new device with the same part number.It was not necessary to switch the surgical technique or do a second surgery.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: d9, g3, h3, h6.The complaint allegation was confirmed.One unpackaged ar-1204af-80 batch 4290153549 was received for evaluation.Visual inspection revealed that the cutting blades were detached from the tip of the shaft.Functional testing was attempted by moving the inner and outer shafts; however, due to the damage, no movement was possible after pressing the button.The most likely causes of the reported failure are misaligned insertion and prying/leveraging the device during insertion.
|
| |
|