| Model Number |
8404BX |
| Medical Device Problem Codes |
Device Reprocessing Problem (1091); Optical Problem (3001)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the device had dimmed light.No patient or procedure involvement.No negative impact in state of health reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The device is pending receipt by manufacturer.The investigation is not yet completed.Investigation results are pending.This event is filed under internal complaint id: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Device received by manufacturer as noted in section d9.The investigation is not yet completed, investigation results are pending.This event is filed under internal complaint id: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Per evaluation from the manufacturer: the errors described by the customer have been confirmed.The device passed the quality check at the time of delivery.Based on the defects found during the technical inspection it is concluded that the most likely cause for the described error is the wear of the adhesive at the tip of the c-mac blade, coming from wear and tear during use and the reprocessing process.The wear of the adhesive causes liquid to penetrate the blade, which affects the electronics, and the light is dim.This event is filed under internal complaint id: (b)(4).
|
| |
|