Catalog Number 3005S3EX
Medical Device Problem Code Device Fell (4014)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This record is a consolidation of records summarized as part of the fda voluntary malfunction summary reporting program.3 devices were functionally/visually inspected in the field.The devices were repaired and returned to use.1 device was not evaluated as the cause was determined by analyzing data provided by the user/third party; no further assistance was requested by the customer.There was no remedial action taken.This device is not labeled for single use.
 
Event or Problem Description
This report summarizes 4 malfunction events, where it was reported the devices experienced fowler unexpected drop/collapse or siderail unexpected drop/collapse.No adverse consequence or clinically relevant delay in treatment was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S3 EX - 3005
Common Device Name
BED, AC-POWERED ADJUSTABLE HOSPITAL
Manufacturer (Section D)
STRYKER MEDICAL-KALAMAZOO
3800 east centre avenue
portage MI 49002
Manufacturer (Section G)
STRYKER MEDICAL-KALAMAZOO
3800 east centre avenue
portage MI 49002
Manufacturer Contact
melissa simon
3800 east centre avenue
portage, MI 49002
2693292100
MDR Report Key21919390
Report Number0001831750-2025-00481
Device Sequence Number9086891
Product Code FNL
UDI-Device Identifier07613327278309
UDI-Public07613327278309
Combination Product (Y/N)N
Initial Reporter CountryUS
Exemption NumberVMSR
Number of Events Summarized4
Summary Report (Y/N)Y
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 04/28/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue Number3005S3EX
Was Device Available for Evaluation? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/01/2025
Initial Report FDA Received Date04/28/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-