ETHICON INC. VCL+ VIO 36IN 1 S/A CTX; SUTURE, ABSORBABLE, SYNTHETIC Back to Search Results
Catalog Number VCP371H
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/31/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a patient underwent a c-section procedure on (b)(6) 2025 and suture was used.During procedure when using suture, the tip of the needle was fractured and it was not known if it remained inside the patient.X-ray equipment was used to look for it but was not found.There were no patient consequences reported.Additional information was requested.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Date sent to the fda: 4/28/2025.H6 component code: g07002 - device not returned this report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent: - was the needle piece(s) retrieved during the same procedure? - was there any additional tissue damage resulting from the search for the needle piece? - does a fragment of the needle remain in the patient¿s tissue? if so, is there any plan in place to remove the needle piece in the future? if so, could you please provide the scheduled date for the removal? in which tissue structure was the broken needle located? what is the surgeon's opinion of the potential consequences for the patient? - were there any unexpected outcomes or complications resulting from the prolonged surgery time? - which tissue was being sutured when the event occurred? - was additional dissection required in other organs/tissues other than the target tissue? - was there any change in the patient¿s post operative care due to the prolonged procedure? - could you kindly perform and document the follow-up attempt for the product return? furthermore, please ensure that the shipment tracking number is recorded in the notes or rmao sections.- please provide the source or name of person providing answers to follow-up questions: the following information was reported: was surgery delayed due to the reported event? yes, if yes, number of minutes: 30, action taken when event occurred? x-ray, was procedure successfully completed? yes, were fragments generated? yes, if yes, were they removed easily without additional intervention? no, if no, explain: the fragment was not found when using x-rays, patient status/ outcome / consequences no, patient consequence description/was there a clinical outcome experienced by the patient (infection, inflammation, etc.)? the patient was discharged, was other medical intervention (e.G.X-rays, additional procedures, prescriptions, otc, revision) required: yes, if yes, describe x-ray, is the patient part of a clinical study unknown to date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).H6 component code: g07002 - no device problem found.This report is being submitted pursuant to the provisions of 21 cfr, part 803.Part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional information: h6, d9, h3.H3 investigation summary: the product was returned to ethicon for evaluation.One open box with eight unopened samples were received for analysis.Product code vcp371h.In order to evaluate the condition of the returned samples, the packets were opened.The swage and attachment area were noted as expected.The tip and body of the needles were examined under magnification and no anomalies or issues related to needle breakage were found.In addition, the samples were tested by resistance test to needle and met the requirements.As part of the ethicon quality process, all devices are manufactured, inspected, and released to approved specifications.Although no conclusion could be reached on the cause of the reported event, the instructions for use do contain the following caution: to avoid this kind of damage: grasp the needle in an area one-third (1/3) to one-half (1/2) of the distance from the attachment end to the point.The event described could not be confirmed as the device performed without any difficulties noted.Although no product defect was identified, there may have been other circumstances or issues that occurred during the use of the device that could not be replicated during the laboratory analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VCL+ VIO 36IN 1 S/A CTX
Common Device Name
SUTURE, ABSORBABLE, SYNTHETIC
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON INC.-BRAZIL
rodovia presidente dutra
km 154
sao paulo 12240
BR   12240
Manufacturer Contact
orla o'mahony
1000 route 202
raritan, NJ 08869
3035526892
MDR Report Key21919676
Report Number2210968-2025-04586
Device Sequence Number9096505
Product Code GAM
UDI-Device Identifier10705031052857
UDI-Public(01)10705031052857(17)300121(10)AV3347
Combination Product (Y/N)Y
Initial Reporter CountryCI
PMA/510(K) Number
K132580
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/30/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberVCP371H
Device Lot NumberAV3347
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/31/2025
Supplement Date Received by Manufacturer05/05/2025
Initial Report FDA Received Date04/28/2025
Supplement Report FDA Received Date05/30/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/11/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-