MAUDE Adverse Event Report: STRYKER MEDICAL-KALAMAZOO PROCUITY ZX WITH ISOTOUR PREM; BED, AC-POWERED ADJUSTABLE HOSPITAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER MEDICAL-KALAMAZOO PROCUITY ZX WITH ISOTOUR PREM; BED, AC-POWERED ADJUSTABLE HOSPITAL Back to Search Results
Catalog Number 3009PX-ZX-450
Medical Device Problem Code Electrical /Electronic Property Problem (1198)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This record is a consolidation of records summarized as part of the fda voluntary malfunction summary reporting program.1 device was not evaluated, as a probable cause for the issue was identified during communication between the customer and stryker and no further assistance was requested by the customer.There was no remedial action taken.This device is not labeled for single use.
 
Event or Problem Description
This report summarizes 1 malfunction event, where it was reported the device experienced accessible ac current.No adverse consequence or clinically relevant delay in treatment was reported.
 
Additional Manufacturer Narrative
In accordance with the ¿final guidance on medical device reporting for manufacturers¿ issued on november 7, 2016, stryker medical will no longer report the hazard of accessible ac current/exposed, bare wires for our bed lines (product code fnl), as this hazard has not caused or contributed to any serious injuries in at least two years.The last serious adverse event manufacturer report #0001831750-2025-00526 was originally submitted on march 2nd, 2023.While no new mdrs will be generated for this hazard moving forward, additional supplemental records may be submitted for past reported events if new information becomes available.Should a new serious adverse event occur attributed to this hazard, mdr reporting will resume.
 
Event or Problem Description
No new information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROCUITY ZX WITH ISOTOUR PREM
Common Device Name
BED, AC-POWERED ADJUSTABLE HOSPITAL
Manufacturer (Section D)
STRYKER MEDICAL-KALAMAZOO
3800 east centre avenue
portage MI 49002
Manufacturer (Section G)
STRYKER MEDICAL-KALAMAZOO
3800 east centre avenue
portage MI 49002
Manufacturer Contact
melissa simon
3800 east centre avenue
portage, MI 49002
2693292100
MDR Report Key21919822
Report Number0001831750-2025-00526
Device Sequence Number8651041
Product Code FNL
UDI-Device Identifier07613327541366
UDI-Public07613327541366
Combination Product (Y/N)N
Initial Reporter CountryUS
Exemption NumberVMSR
Number of Events Summarized1
Summary Report (Y/N)Y
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/29/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue Number3009PX-ZX-450
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/01/2025
Supplement Date Received by Manufacturer04/01/2025
Initial Report FDA Received Date04/28/2025
Supplement Report FDA Received Date04/29/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-