Additional Manufacturer Narrative
|
|
This record is a consolidation of records summarized as part of the fda voluntary malfunction summary reporting program.6 devices were functionally/visually inspected in the field.The devices were repaired and returned to use.3 devices are pending evaluation.There was no remedial action taken.This device is not labeled for single use.
|
| |
|
Event or Problem Description
|
|
This report summarizes 9 malfunction events, where it was reported the devices experienced reduced/inadequate brake force.No adverse consequence or clinically relevant delay in treatment was reported.
|
| |
|
Event or Problem Description
|
|
No new information.
|
| |
|
Additional Manufacturer Narrative
|
|
2 devices that were pending evaluation were evaluated in the field and the issue was confirmed.The devices were repaired on site and returned to service.1 device that was pending evaluation was not made available by the customer; the reported issue was not confirmed.Section h codes have been updated to reflect this.
|
| |
|