Catalog Number VCPP81D
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/27/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).H3 investigation summary: this is an analysis for a photo submitted to ethicon for evaluation.During the visual analysis, the following was observed: the photos show an open foil, and a winding with eight strand product code vcpp81d was observed, with five needles without sutures.A clear analysis of the end point of failure or the needle hole could not be performed due to the position of the needle.Based on the photo review, the event describe is confirmed, however no conclusion or root cause could be determined.Hands on device analysis may provide the additional evidence necessary to confirm the root cause of the reported event.Because the device was not returned our evaluation is limited.We value the opportunity to fully analyze the device upon its return.As part of ethicon quality process all devices are manufactured, inspected, and released to approved specifications.H6 component code: g07002 - device not returned.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.The following information was requested, but unavailable: were there any patient consequences? a photo with 8 needles (3 attached to a suture and 5 without) was provided.Please confirm the number of devices in which the needle "pull off" during this procedure.Received information as following from sales rep via phone today.Please refer to the event description, other information requested is unknown.
 
Event or Problem Description
It was reported that a patient underwent an unknown procedure on (b)(6) 2025 and suture was used.During the procedure, when suturing the patient with suture, it was found that the needle and suture connection was loose and easy to break.Stop using it immediately.There were no adverse reported.Additional information was requested.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VCL+ VIO 8X27IN 0 S/A MO-5 CR
Common Device Name
SUTURE, ABSORBABLE, SYNTHETIC
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON INC.-JUAREZ
calle durango #2751
ciudad juarez 32575
MX   32575
Manufacturer Contact
orla o'mahony
1000 route 202
raritan, NJ 08869
3035526892
MDR Report Key21920854
Report Number2210968-2025-04605
Device Sequence Number8669534
Product Code GAM
UDI-Device Identifier10705031056886
UDI-Public(01)10705031056886(17)290228(10)UCMCHU
Combination Product (Y/N)Y
Initial Reporter CountryCH
PMA/510(K) Number
K032420
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 04/28/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberVCPP81D
Device Lot NumberUCMCHU
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/31/2025
Initial Report FDA Received Date04/28/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured03/07/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-