• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: JABIL CIRCUIT SINGAPORE PTE LTD BIS; INDEX-GENERATING ELECTROENCEPHALOGRAPH SOFTWAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

JABIL CIRCUIT SINGAPORE PTE LTD BIS; INDEX-GENERATING ELECTROENCEPHALOGRAPH SOFTWAR Back to Search Results
Model Number PMB4000-CE
Medical Device Problem Code Incorrect Measurement (1383)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/15/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
It was reported that the device had an inaccurate sensor check value during testing and no baseline need to be set for test sensor.There was no patient involvement.
 
Additional Manufacturer Narrative
Additional information: g3, h3, h6.Evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.The evaluation found no potentially contributing factors, and the sample met all related specifications.It was reported that the device had an inaccurate sensor check value.The reported issue could not be confirmed.The most likely cause could not be identified because no related problem was detected with the device.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIS
Common Device Name
INDEX-GENERATING ELECTROENCEPHALOGRAPH SOFTWAR
Manufacturer (Section D)
JABIL CIRCUIT SINGAPORE PTE LTD
16 tampines crescent
singapore 52860 4
SN  528604
Manufacturer (Section G)
JABIL CIRCUIT SINGAPORE PTE LTD
16 tampines crescent
singapore 52860 4
SN   528604
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key21922630
Report Number2936999-2025-00223
Device Sequence Number8650624
Product Code OLW
UDI-Device Identifier10884521808270
UDI-Public10884521808270
Combination Product (Y/N)N
Initial Reporter CountryHK
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/28/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberPMB4000-CE
Device Catalogue NumberPMB4000-CE
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/16/2025
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/02/2025
Supplement Date Received by Manufacturer05/26/2025
Initial Report FDA Received Date04/28/2025
Supplement Report FDA Received Date05/28/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/14/2024
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-