SMITH & NEPHEW, INC. LEGION 6 DEG VALGUS COLLET 6MM OFFSET SZ3-8; PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
|
Back to Search Results |
|
| Catalog Number |
71434407 |
| Medical Device Problem Codes |
Material Fragmentation (1261); Material Deformation (2976); Positioning Problem (3009)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/18/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Internal complaint reference: (b)(4).The associated device was returned and evaluated.A visual inspection of the returned device concluded the device presents with significant damage to edges and surfaces.There are several areas where the metal has been chipped, gouged and deformed along the edges of the offset cllet inner taper 6mm and the base of the offset collet knob 6mm.The device is unstable.The damage found indicates that the device will not function as intended.A review of the production order did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history for the part number over the past 12 months and for the batch number based on historical data of the device did not reveal similar events for the listed device.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no evidence to conclude that the product failed to meet any specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely potential factors that could contribute to the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|
| |
|
Event or Problem Description
|
|
It was reported that, during a tka surgery, one (1) legion 6 deg valgus collet 6mm offset sz3-8 would not seat around the revision reamer.The procedure was resumed, after a delay of less than 30 minutes, using a different surgical technique.No injury was reported as a consequence of this issue.
|
| |
|
Search Alerts/Recalls
|
|
|