| Catalog Number |
10-240 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
04/01/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported right side rupture and anxiety.Device has been explanted and replaced.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported right side rupture and anxiety.Device has been explanted and replaced.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: based on the product analysis performed, the assessments of the complaints are: rupture: observed, opening assessed as fold flaw opening.Anxiety - product/ procedure: unable to observe since it is a medical event and is not related to the device.As per the investigation procedure, crease, wear abrasion and non-penetrating nick was completed and none of the observations are found to be potentially related to the manufacturing process, no further actions are required.Additional, changed, and/or corrected data: d9; h3; h6.
|
| |
|