The manufacturer was contacted in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.The manufacturer received information alleging device won't turn on/device not functioning and particles in airpath.During the evaluation of the device at the third-party service center, the device was visually inspected and no visible foam degradation was observed.The device's event log was reviewed by the service center and no error code found.Additionally, the device was scrapped.There was no allegation of serious or permanent harm or injury.No medical intervention was specified by the patient.The manufacturer was made aware of this complaint through a representative of the customer.At this time, no further investigation can be performed.If any additional information is received, a follow up report will be filed.
|