| Catalog Number |
111653 |
| Medical Device Problem Codes |
Fracture (1260); Material Fragmentation (1261)
|
| Health Effect - Clinical Codes |
Foreign Body In Patient (2687); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/07/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
As part of normal complaint follow-up, an evaluation of the event has been initiated by mako surgical.A supplemental report will be submitted if additional information becomes available.
|
| |
|
Event or Problem Description
|
|
The sales rep reported: "checkpoint (111653) 3.5 hex tip broke whilst removing it from the acetabular bone.".
|
| |
|
Event or Problem Description
|
|
The sales rep reported: "checkpoint (b)(6) 3.5 hex tip broke whilst removing it from the acetabular bone." update: the surgeon was unable to retrieve the tip of the checkpoint.
|
| |
|
Additional Manufacturer Narrative
|
|
An event regarding crack/fracture involving a mako checkpoint was reported.The event was not confirmed because the product was not available for inspection.Method & results: -product evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device was not returned.-clinician review: no medical records were received for review with a clinical consultant.-product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.-complaint history review: there have been other complaints with similar event(s) for the lot referenced.Conclusions: the alleged failure mode was not confirmed because the product was not available for inspection.No additional investigation or specific actions are required.If additional information is received then the complaint will be reopened.
|
| |
|
Search Alerts/Recalls
|